1.1 General Information
Digital Compactibility Tester is equipped with a squeezing mechanism and sensing mechanism which are controlled by Microcontroller or PLC and observed on LCD display.
1.2 General Usage
To check compactibility & to prepare standard specimen under identical condition for checking important physical properties of moulding sand.
1.3 Product Data
Product Name:
Digital Compctibility Tester
Product Code:
VCCD
Order Reference Code:
V9001343
Electrical Supply Required:
230v 50Hz A.C. (Default)
110v 60 Hz A.C (Optional)
Single Phase, Stabilized with Over/ Under Voltage cut-off.
Electrical Plug:
M type : 15 Amp Plug
Typical Maximum Current:
2.2 A at 230V
Compressed air Supply:
5 Kg/cm2 Dry Air
Communication:
RS485 MODBUS RTU
RS232 Serial
Ethernet (Optional)
CloudSync using V-Sync
Dimensions:
465x300x665 mm (WXBXH)
Packing Dimensions:
565x400x815 mm (WXBXH) Typical
Net Weight:
Approximately 52 kg
Gross Weight:
Approximately 60 kg
Colour:
Factory Default
Measuring Range:
0 to 52 %
Traceable to:
NABL Standards
Least Count:
0.1 %
Specimen Requirements:
Not Required
Operating Temperature:
0-500C
Testing Standards:
BIS: Standard
AFS: Optional
Calibration Frequency:
22500 nos. of Testing Cycles or Once in Year
2.0 Scope of Supply
2.1 Inclusions in Supply Scope
Digital Compatibility Tester.
Specimen tubes, Pedestal cup & Stripper.
Calibration certificate.
Instruction manual.
2.2 Exclusions in Supply Scope
Data interfacing accessories.
Anything that is not covered under Supply scope.
3.0 Recommended Pre-requisite Equipment
Sand Splitter : To get representative sample
Tube Filler Accessory VTF.
Weighing Balance (LC 10mg)
4.0 Accessories
4.1 Calibration Kits
Calibration Piece High VCCD
Calibration Piece Low VCCD
4.2 Attachments
NA
5.0 Alternative Equipment
Sand Squeezer VCC.
6.0 Installations Instructions
Required Levelled (2mm) and sturdy platform. (Equipment should be away from external source of vibration or electromagnetic interference.) Other Installation instruction shall be provided along with the equipment.
7.0 Consumables
8.0 Hazards
Squeezing Hazard: Do not insert hand or limb under the plunger when in operation.
9.0 Additional Information
Link to Web-Page