1.1 General Information
High Dry Strength Attachment is equipped with a hydraulic loading mechanism and it is suitable to connect to Universal strength machine model VUN. It can measure the strength of an AFS (Dia. 2”X2”) or Metric (Dia. 50mmX50mm) Specimen.
1.2 General Usage
Keep the specimen in between compression pads or accessory, start the test cycle & get the various type of strength reading on base equipment.
1.3 Product Data
Product Name:
High Dry Strength Attachment
Product Code:
VUH
Order Reference Code:
V0000062
Electrical Supply Required:
Not Required
Electrical Plug:
Not Applicable
Typical Maximum Current:
Compressed air Supply:
Communication:
Dimensions:
430x290x220mm (WXBXH)
Packing Dimensions:
550x400x400mm (WXBXH) Typical
Net Weight:
Approximately 16.7 kg
Gross Weight:
Approximately 19 kg
Colour:
Factory Default
Measuring Range:
Compression:0-200 Kg/cm2
Shear : 0-150 Kg/cm2
Traceable to:
Manufacturers Standards
Least Count:
Compression:4 Kg/cm2
Shear : 2 Kg/cm2
Specimen Requirements:
Compression type
Operating Temperature:
0-500C
Testing Standards:
BIS: Standard
AFS: Optional
Calibration Frequency:
15000 nos. of Testing Cycles or Once in Year
2.0 Scope of Supply
2.1 Inclusions in Supply Scope
High Dry Strength Attachment.
Assy Funnel.
Working condition certificate.
Instruction manual.
2.2 Exclusions in Supply Scope
Anything that is not covered under Supply scope.
3.0 Recommended Pre-requisite Equipment
Sand Splitter : To get representative sample
Universal Strength Machine VUN
Specimen preparation equipment’s to make a standard AFS or Metric specimen.
4.0 Accessories
4.1 Calibration Kits
NA
4.2 Attachments
VHS: Shear Strength Attachment
5.0 Alternative Equipment
6.0 Installations Instructions
Required Levelled (2mm) and sturdy platform. (Equipment should be away from external source of vibration) Other Installation instruction shall be provided along with the equipment.
7.0 Consumables
Hydraulic oil 320 No. # V4000182
8.0 Hazards
Squeezing Hazard: Do not insert hand or limb in between the pads when in operation.
9.0 Additional Information
Link to Web-Page
Disclaimer: Specifications and design of equipment may change without prior notice.